Sterilization and de-pyrogenation Tunnel |
Sterilization and de-pyrogenation
Sterilization Tunnel used to destroy any microorganisms by
heat for the empty washed pharmaceutical vials to be ready for safe product packing.
The contamination level of the pharmaceutical vials depends
on the heating exposure (more than 300◦C) and exposure duration “cycle time” which,
have not adverse effects on the product parameters.
Sterilization Tunnel Zones
1- In-feed Zone
Consisting of Tunnel conveyor belt, Tunnel conveyor belt
Drive and Laminar flow in-feed zone “H.E.P.A Absolute filters, Absolute filter
Differential gauge, Pre-filter and Pre-filter Differential gauge”.
The vials coming from the upstream machine (washing machine)
to the tunnel in-feed zone then the tunnel conveyor belt conveying the vials to
the central irradiation heating zone.
2- Central Irradiation Heating Zone
Consisting of heating Chamber, Side doors, Stainless steel
bellows (To totally prevent any release of particles from or to the
sterilization zone), Top Irradiation lamps, Lower Irradiation lamps, spring
supports, Modulating heat-regulation system, temperature probes, Additional
temperature probe, Ceramic Insulation, Fiberglass Insulation, Transmitters, Heat
regulators and exit hot chamber shutter.
The tunnel belt moving synchronized with the upstream
machine and when the vials reached the first external shutter lifted
efficiently approximately 5 cm from the first shutter then the shutter lifted
by a lever and locked in a position proper for the vials size. After "cycle time" the vials reached the hot chamber discharge
shutter which, opened and locked to allow vials conveying to the cooling zone.
3- Cooling Zone
Consisting of laminar flow, cooling batteries, absolute
filters, Lower plenum, Hot wire anemometer, Differential gauge (For visual
check of pressure drop at absolute filter), Delivery hood, Manual adjusting
vertical shutter, heat regulator, cooling battery and exit cold chamber shutter.
Air pressure difference between in-feed and cooling chamber must be automatically regulated to prevent the possibility of counter air flow from cooling chamber to the in-feed chamber. Then vials conveyed to the down stream machine (Filling machine) through the rotary table.
Standard Instrumentation and controls
Consisting of control panel (Lockable stainless
steel panel), Power supply, Output relays, Proximity sensors, temperature
probes, Potentiometric recorder, PLC and Process parameter recorder.